SYSTEM TROUBLESHOOTING AND ALARMS

PROVIDER TROUBLESHOOTING TABLE
Symptom
On Start Up (first 5 seconds)
- unit beeps 3 times
On Start Up (first 5 seconds)
- unit does not beep at all
Symptom
Operating Any
Power Source: Yel-
low Light is "blinking"
(Low O2 production 85 - 70%)
Alarm is silent
Press NO SMOKING
button for Diagnostic Code
"ALRM=008"
Operating on Battery
Power: Yellow Light is "blink-
ing" and unit is "beeping"
Press NO SMOKING
button for Diagnostic Code
"ALRM=010"
Any Power Source:
Yellow Light is "Blink-
ing" and Unit is "beeping"
(Low or No Oxygen Flow)
Press NO SMOKING
button for Diagnostic Code
"ALRM=020"
On Battery Power:
Yellow light is "On" and
"Steady" - Alarm is silent
Warm Battery
Press NO SMOKING
button for Diagnostic Code
"ALRM=002"
Personal Ambulatory Oxygen System Provider Technical Manual
DO NOT IGNORE ALARMS.
Probable Cause
The 9V battery (internal to the Eclipse)
is low
The 9V battery (internal to the Eclipse)
is below specification
YELLOW LIGHT
Probable Cause
Restricted Air Flow through cabinet
inlet filter
Blocked Air Flow to cabinet inlet filter Ensure the Eclipse has adequate air flow around the device (6"
Restricted Air Flow through Internal
Filters
Pulse Mode: Out Breathing the Eclipse
exceeding Breaths Per Minute for that
Pulse setting
Low Battery: Running on Battery
power, not plugged into external
power
Loss of External Power: Plugged
in to external power but running on
Internal Battery
Battery is out of calibration
Eclipse is ON . . . Filter Blocked
Humidifier attached?
Tubing is not properly connected or
kinked
Other
Warm Battery Condition: The Battery's
internal temperature has increased to
a point above 113F (45C) degrees that
will not allow the Battery to charge.
When in use, above 140F (60C) the
battery will flag the code 002, and shut
down about 158F (70C).
Shut Down due to overheating: the
Eclipse displays a FAIL 92, FAIL 91
Remedy
Replace 9V battery/Perform Annual Preventive Maintenance (5022-
SEQ)
Replace 9V battery/Perform Annual Preventive Maintenance (5022-
SEQ)
Remedy
Perform Annual Preventive Maintenance (5022-SEQ); if issue remains
call Chart TS for service
around the device) - turn the device around in the car(exhaust fac-
ing to the middle of the car)
Perform Annual Preventive Maintenance (5022-SEQ); if issue remains
call Chart TS for service
Patient's breathing is too fast for Breath Rate to operate within that
Pulse setting's specifications(See table for max BR in Pulse.). Can be
verified using the EDAT software program (PN: 5535-SEQ)
During Battery use, the Battery the indicator will decrease until a
Low Battery condition is met. Change with a fresh battery.
Verify the Power Verification Indicators (Green LEDs) on the power
supply and the membrane panel are ON
Unplug the Eclipse from EXT power, Run the battery all the way
down until the Eclipse shuts down and alarms, Plug it in to AC
power and allow to recharge fully. Replace battery, if it does not
come back within specification.
Perform Annual Preventive Maintenance (5022-SEQ), if issue remains
call Chart TS for service.
*Check/Clean Humidifier and tubing
*Remove Humidifier if in Pulse Mode
Check Tubing, Humidifier and Cannula for blockage, and clean as
needed
Contact Chart TS for service
Allow the Eclipse and Battery to Cool for 30 minutes Minimum.
(Remove battery and store in a cool place, re-insert the battery.)
Replace with a fresh cool Battery
Reset Power to the Eclipse. (Remove Battery and External Power
together, wait 30 minutes or until cool. Re-insert External power
or re-insert the battery.) If condition remains contact Chart TS for
service