Assembly And Alarm Verification Tests - CAIRE 1000B Provider Technical Maunal

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

ASSEMBLY AND ALARM VERIFICATION TESTS

To ensure proper assembly and functionality of the Eclipse after it has been reassembled, the following steps should be
followed.
1. Install the Power Cartridge into the Power Cartridge compartment of the Eclipse. Plug the AC Power Supply int the
wall outlet and connect it to the External Power Connector of the Eclipse.
2. Press the ON button and set the Eclipse to 2 LPM. At initial start-up Eclipse units light the Green, Yellow and Red LED's
accompanied by a beep. As concentration increases to ~70%, the Red LED turns off and the Yellow and Green remain
on. When the concentration reaches normal operating range (above 85%), the Yellow LED will turn off. The Green LED
will stay lit indicating normal operation. Table 3 shows the normal start-up operating conditions for Eclipse Oxygen
Systems. If LED is not green within 7 minutes there may be a problem with the Eclipse.
Green
Indicator
Off
Off
On
3. Verify that the Power Cartridge is charging as indicated by the Power Cartridge Status Gauge is moving in waterfall
fashion (scrolling from bottom to top). If the Power Cartridge Status Gauge is not moving, verify that the Power
Cartridge is properly engaged.
4. Check the Control Panel by pressing each of the Increase/Decrease, Pulse Dose Mode and No Smoking buttons and
observe that the Eclipse buttons function normally.
5. Press the Delivery Mode button and set the Eclipse to the patient's normal Pulse Dose setting. Confirm that without
breathing from the unit, after 15 seconds of no breath detected, the system changes automatically to Continuous
Flow Mode. The green Delivery Mode light is blinking fast, indicating the system is delivering a Continuous Flow. After
another 15 seconds, the system stops delivering Continuous Flow and waits 15 seconds trying to detect a breath. The
cycle continues until a breath is detected.
6. After 5 minutes of operation, block the Cannula Fitting Outlet with your finger for 2 minutes and confirm that the
visual and audible alarm occurs. Unblock the Cannula Outlet Port and let it stabilize until the alarm stops.
7. Disconnect the AC Power Supply and allow the Eclipse to run for about one minute. Set the Eclipse to the patient's
normal Continuous Mode setting. Confirm that the Power Cartridge Status Gauge is illuminated and External Power
Present Indicator is off.
8. Remove the Power Cartridge and verify that the Eclipse alarms and red Alarm Indicator is flashing and the alarm
sounds. After about 5 seconds, re-install the Power Cartridge and connect the AC Adapter and observe that the
Eclipse automatically restarts. Confirm that the External Power Present Indicator is illuminated.
9. Turn off Eclipse and unplug the AC Power Supply.
10. Record results, initial and date the Service and Maintenance Record.
11. EDAT may be utilized to record Eclipse device status and dates of service.
Personal Ambulatory Oxygen System Provider Technical Manual
Yellow
Red Indicator
Indicator
Off
Blinking
Blinking
Off
Off
Off
Audible
Alarm code
Alarm
Beeps
004
Off
008
Off
000
Operating Condition
Purity < 70%
Purity between 70% and 85%
Normal Operating Condition
51

Advertisement

Table of Contents
loading

Table of Contents