between portable and mobile RF communications equipment and the Eclipse 3
The Eclipse 3 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The cus-
tomer or the user of the Eclipse 3 can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Eclipse 3 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum out-
put power of transmit-
ter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Personal Ambulatory Oxygen System Provider Technical Manual
Recommended separation distances
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
P
1
2 .
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
800 MHz to 2,5 GHz
d
P
2
3 .
0.23
0.73
2.3
7.3
23