ALARM CONDITIONS AND ALARM CODES

Use the table below to decode Eclipse alarm conditions. If other alarm codes are displayed by the Eclipse, contact Chart-
SeQual Technical Service for assistance.
Note: The following table is intended as a guide for the provider, not the user.
ALARM
CONDITION
CODE
"Warming Up. Please
004
Wait. "
"Warming Up. Please Wait. "
008
"3.0" (or flow setting)
000
"Low Power Cartridge"
010
"Warm Power Cartridge"
002
"Low 9V
-
Battery"
"O
Concentration < 85%"
008
2
"O
Concentration < 70%"
004
2
"Flow Rate Error / Blocked
020
Flow"
"Cannot Charge Power
001
Cartridge"
"One Wire Communica-
100
tion Loss"
"No Inspiration Detected"
200
(in Pulse Mode only)
"Loss of External Power"
(without charged Power
040
Cartridge installed)
"System Fault"
080
Personal Ambulatory Oxygen System Provider Technical Manual
ALARM
GREEN
YELLOW
DISPLAY
INDICATOR
INDICATOR
O2 < 70%
ON
O2 < 85%
ON
NO ALARM
ON
LOW BAT
ON
Flashing
WARM BAT
ON
Flashing
-
OFF
O2 < 85%
OFF
Flashing
O2 < 70%
OFF
FLOWRATE
OFF
Flashing
CHARGER
N/A
ONE WIRE
N/A
Flashing
P <--> C
ON
********
OFF
(flashing)
FAIL XX
OFF
RED
AUDIBLE
INDICATOR
ALARM
ON
ON
OFF
ON
OFF
OFF
OFF
OFF
OFF
OFF
2 Beeps
OFF
1 Beep
3 Beeps on
OFF
OFF
POST
OFF
OFF
OFF
Flashing
3 Beeps
OFF
1 Beep
ON
OFF
OFF
OFF
1 Beep
OFF
OFF
OFF
OFF
Flashing
ON for 5 min
ON 10 seconds
OFF
ON
then silent
WHAT TO DO
Wait. The system typically takes 3 – 5 minutes to reach
specified performance.
Wait. The system typically takes 3 – 5 minutes to reach
specified performance.
Nothing. The system is operating properly at the specified
flow rate.
Plug into external power or replace with a charged Power
Cartridge.
Plug into external power or replace with a charged Power
Cartridge.
Replace the internal 9-volt battery.
Perform Annual Filter PM Maintenance. If condition persists,
service is required.
Perform Annual Filter PM Maintenance. If condition persists,
service is required.
Check tubing and/or humidifier for obstruction. Check/clean
air intake filter. If condition persists past 10 minutes, service
is required.
Instruct patient to seek an external power source. Remove
Power Cartridge and allow to cool to room temperature. If
Power Cartridge malfunction persists, service is required.
Check battery connections.
If no inspiratory effort is detected after 15 seconds, the
Eclipse will switch to Continuous Flow for 15 seconds, and
continue this sequence until a breath is detected.
Instruct patient to seek an alternative oxygen supply
until external power is restored or install a charged Power
Cartridge.
Reset power, wait 20 seconds, reconnect power, restart
device.
Instruct patient to use back up oxygen supply. Service is
required.
60