Shipping And Transporting The Eclipse; Storing The Eclipse; Discarding - CAIRE 1000B Provider Technical Maunal

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

SHIPPING AND TRANSPORTING THE ECLIPSE

When shipping the Eclipse use original packaging, if possible. Always remove the Power Cartridge and cart from the Eclipse
prior to shipping.
If original packaging material is available repack the Eclipse, Power Cartridge, cart and power supplies in the designated
packaging areas.
If original packaging material is not available, then place the Eclipse in a plastic bag and surround the concentrator with a
minimum of two inches of soft foam packing material or bubble wrap. Wrap each accessory in a similar manner. Place the
Eclipse and accessories in an appropriate cardboard box for shipping.
WARNING:
When the Eclipse must be transported in a delivery vehicle, simply secure the Eclipse and its accessories to prevent damage.
Do not expose the device to extreme heat, cold, or humidity.

STORING THE ECLIPSE

Heat and humidity may degrade performance or severely damage the Eclipse. Store the device in a cool, dry, protected area
away from high temperatures, moisture and humidity. Remove the Power Cartridge when storing the device.
The Eclipse must be turned on and run for 2 hours each month to ensure proper operation performance.
Tip: Running the power cartridge down completely and recharging it will provide two different features for the Eclipse:
1. Ensures the battery cartridge is calibrated
2. Running the Eclipse between uses while in storage allows the user to know if service may be needed prior to
needing the device.

DISCARDING

Power Cartridge Disposal:
the local city or town offices for instructions on proper disposal of the Power Cartridge. Alternately, Chart SeQual may be
contacted for the Power Cartridge disposal.
Eclipse Oxygen System Disposal:
equipment such as the Eclipse, AC Power Supply or the DC Power Supply. Contact the local city or town offices for instructions
on proper disposal of electrical or electronic equipment. Alternately, Chart SeQual may be contacted for disposal information.
Personal Ambulatory Oxygen System Provider Technical Manual
DO NOT expose the Eclipse to water. The Eclipse enclosure does
not provide protection against the harmful effects of liquid ingress.
Electrical shock or damage to the unit may result.
The Power Cartridge can be recycled and should not be thrown into the trash. Contact
Local environmental laws may prohibit disposal of electrical and/ or electronic
55

Advertisement

Table of Contents
loading

Table of Contents