SHOWING POWER CARTRIDGE POWER LEVEL

The display on the Control Panel shows the amount of Power Cartridge capacity available and waterfalls when charging. Point
out the table showing typical new Power Cartridge duration-of-use time in the Users Manual.
WARNING:

SELECTING THE PROPER FLOW MODE AND SETTING

The recommendations of the qualified clinician responsible for the patient's care should always be followed when instructing
patients on the use of Continuous or Pulse Dose Mode.
WARNING:

SELECTING CONTINUOUS FLOW MODE

When delivered from SeQual, the Eclipse is set to Continuous Flow Mode. Refer your patient to the Users Manual for instructions
on how to select Continuous Flow Mode. Demonstrate the procedure on the device. Set the prescribed flow using the increase
or decrease flow rate buttons
CAUTION

SELECTING PULSE DOSE MODE

Pulse Dose Mode, delivers a measured bolus of oxygen pulse at the beginning of each inspiration. Refer to the Users
Manual for instructions on how to select Pulse Dose Mode. Demonstrate the procedure for your patient. Set the prescribed
flow using the increase or decrease buttons. Verify that the patient is able to adequately trigger the oxygen delivery while
speaking, sitting quietly, and walking. If the patient is not able to satisfactorily trigger oxygen delivery, the Pulse Sensitivity
may need to be adjusted.
CAUTION
Review with the patient the breath rate capabilities of Pulse Dose Mode for their prescription.
Personal Ambulatory Oxygen System Provider Technical Manual
The display gives an approximate level of remaining battery power.
If the patient is dependent on oxygen therapy, the users should
always have a backup oxygen supply or access to AC or DC Power
Supply in the event of loss of battery power.
Do not attempt to prescribe oxygen level settings. Warn your patient
that only a qualified clinician or physician is qualified to perform the
necessary tests to prescribe treatment.
• Continuous Flow Mode is required in order to use a humidifier bottle.
• Continuous Flow Mode without a humidifier bottle is required in order
use 50 foot tubing.
• Do not use a humidifier when in Pulse Dose Mode.
• Use only a 7-foot cannula when in Pulse Dose Mode. Do not use
a tubing extension.
See Table on page 7