Annual Maintenance—Provider

INTRODUCTION

Properly maintaining the Eclipse will ensure longer life and higher performance. Minimum annual maintenance is required.
CAUTION

ANNUAL MAINTENANCE CHECK LIST

Perform the following maintenance procedures at least once a year or more often., as needed. The frequency of the periodic
maintenance should be based on the environment in which the Eclipse is used.
• Replace air inlet filter
• Check Power Cartridge
• Inspect AC and DC Power Supply plugs and cords for damage. Replace as needed
• Verify that the AC and DC Power Supplies operate with the Eclipse
• Readjust the Pulse Dose Sensitivity setting between patients as needed
• Readjust the Rise Time setting between patients, as needed
• Replace 9V battery
• Check Universal Cart for functioning wheels and telescoping handle. Replace as needed
• Read and record hour meter
• Check flow rate, purity, Pulse Dose Mode and alarm functions
• Replace compressor intake filter
• Perform electrical safety test (Required only for PN 5701, Hospital Grade Power Cord)
• Replace HEPA filter

CHECKING AND REPLACING THE POWER CARTRIDGE

Typically, the Power Cartridge will require no routine maintenance beyond cleaning by the patient and calibration. SeQual
recommends that the Power Cartridge be replaced if there is physical damage to the cartridge enclosure or connector, or if the
battery's recharge capacity, after calibration, is 20% less than published values at its given flow rate.
ECLIPSE MONTHLY RUN-TIME PROCEDURE
1. Power on Eclipse using AC Power, DC Power, or Power Cartridge
2. Allow unit to run for a minimum of 2 hours
This procedure is performed to ensure and prolong the life of your Eclipse. This is also an ample
NOTE:
opportunity to calibrate the power cartridge of the Eclipse.
Personal Ambulatory Oxygen System Provider Technical Manual
The Eclipse contains electrostatic sensitive components. Do not
open or handle except at a static free workstation. Do not remove
cover without ESD protection.