CAIRE 1000B Provider Technical Maunal page 33

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

CAUTION
WARNING:
Use the following table to determine whether your patient's prescribed flow setting is supported by the Eclipse while on DC
power.
FLOW SETTINGS AVAILABLE WHEN OPERATING ON DC
Flow Setting
0.5
1.0
1.5
2.0
2.5
3.0
-
-
-
-
-
-
-
-
-
NOTE: Actual performance of the Eclipse may vary depending upon usage conditions of the motor vehicle.
CAUTION
Personal Ambulatory Oxygen System Provider Technical Manual
Be sure to accurately determine the amount of current the vehicle
accessory outlet is rated to supply.
Avoid placing the Eclipse in direct sunlight.
Do not store the Eclipse in a vehicle where the device may be subject to extreme
temperatures. Extreme heat or cold may impair operation and damage the device
and degrade the Power Cartridge.
When using the Eclipse in a vehicle, check the cannula to make sure it
is not pinched or occluded.
Continuous
Setting / Bolus Size (mL)
Yes
Yes
Yes
Yes
NO
NO
-
-
-
-
-
-
-
-
-
Flow Modes for DC Power Supply Operation.
When operating with an Eclipse 3 DC Power Supply, Eclipse settings
greater than 2.0 LPM Continuous Flow, will automatically switch to the
2.0 LPM Continuous Flow setting. 2.0 LPM is max. Pulse mode will also
not operate above 96 mL.
Pulse Dose
0 / -
1 / 16
1.5 / 24
2 / 32
2.5 / 40
3 / 48
3.5 / 56
4 / 64
4.5 / 72
5 / 80
5.5 / 88
6 / 96
128
160
192
-
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
NO
NO
NO