Record Hours Of Operation & Software Version - CAIRE 1000B Provider Technical Maunal

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

RECORD HOURS OF OPERATION AND SOFTWARE VERSION
To help maintain the Eclipse, you may obtain the Total Hours of Operation and software version numbers for the Control
Board and the Power Manager Systems by following the steps below.
- If connected to the AC or DC Power Supplies, press
the "No Smoking" icon on the control panel and
advance to the following display mode for each time
the icon is depressed:
• Alarm Code (ALRM) - Displays most recent
alarm code. Additional alarm codes
will also be displayed, if present, by
continuing to press the + key.
• Pulse Sensitivity (PS)
• Hours of Operation (HRS)
• 9-volt Battery Status (9V)
• Rise Time (BOL)
• Control Board Software Revision
• Power Manager Software Revision
- If external power is not available and the power
cartridge is installed, start by pressing the ON/OFF key
for 1 second.
- If no key press is sensed within 5 seconds, the Eclipse
will drop out of Provider Mode.
Personal Ambulatory Oxygen System Provider Technical Manual
Alarm Code
Pulse Sensitivity
Hours of
Operation
9-Volt Battery
Status (Volts)
Rise Time
Control Board
Software Revision
Power Manager
Board Software
Revision
Log the Results
NOTE:
Data Output Form Example
The display accrues to "99,999.9"
hours, then rolls over to "00,000".
52

Advertisement

Table of Contents
loading

Table of Contents