5. Locate the large O-ring on the product tank, replace it with the new large O-Ring included in the PM kit (5022-SEQ). Screw the
new HEPA filter into the carriage holder (wing nut topped housing). Insert and screw the new HEPA filter (wing nut topped housing)
into the Carriage holder (product tank) and finger snug in place (do not over-tighten). Re-attach Tubing and secure with a cable tie.

REMOVE AND REPLACE THE COMPRESSOR INTAKE FILTER

Replace the compressor intake filter annually, or more often as needed. The compressor intake filter may become clogged
depending upon the amount of contaminants in the air (smoke, dust, dirt, pollen, etc.) and may need to be changed more
frequently. The most common cause of low concentration and eventual system failure is a dirty or clogged compressor inlet filter.
Environmental conditions usually determine the effective life of a filter.
CAUTION
NOTE:
Personal Ambulatory Oxygen System Provider Technical Manual
Figure 26 O-Ring.
Operating the Eclipse with a clogged compressor intake filter may
reduce performance and lead to system damage or premature failure.
Always cut the heads of cable ties to avoid damaging the tubing.