REMOVE AND REPLACE THE PRODUCT TANK ASSEMBLY (PN 4378-SEQ)
ECLIPSE 3-2011 IN SN 11F06100456 AND ABOVE
Situation# 1) If you are "upgrading" from a previous model Product Tank (PN1038-SEQ) you will need to let Tech Support know
as the ATF output port is in a different location as it can be different between ATF models (PN: 3995 to PN: 5494). Output
port has changed location. Then follow steps for Situation 1.
Situation# 2) If you are removing the Product Tank for a failure of one of its components then follow steps for Situation 3.
Situation# 3) If you are just replacing the HEPA filter then follow the steps in the PM section for Remove and Replace HEPA
Filter.
Situation# 1: Removal of Old Style Product Tank(PN: 6138-SEQ)
1) Follow the steps to Removing the Unit Cover
2) Remove the Front Cover Assembly completely; outlet tubing and front membrane ribbon cable. Set aside.
3) Locate the Product Tank and cut the cable ties from the ATF to the product tank (Bottom), and O2 out from the outlet
connector (Top). Remove tubing from ATF to lower product tank barb. See Figure 43.
Important: Cap off the ATF hose barb with a black hose barb cap.
Personal Ambulatory Oxygen System Provider Technical Manual
Figure 43: Cut Cable Ties
Cut Cable
Ties