CAIRE 1000B Provider Technical Maunal page 81

Eclipse3 with autosat personal ambulatory oxygen system
Table of Contents

Advertisement

3.
Disconnect the 5 wire harnesses as shown in Figure 49. Remove the Battery Bridge PCB as shown in Figure 57.
External Power
Wire Harness
Blower Wire
Harness
4.
Unscrew the 2 screws (M2x10 Pan Head Machine Screw, P/N 6961-210-SEQ) holding the fan shown in Figure
49. Remove fan.
5.
Remove the Exhaust Tube shown in Figure 58.
Exhaust
Tube
Personal Ambulatory Oxygen System Provider Technical Manual
Compressor Signal
Wire Harness
Figure 57: Power Manager PC Board Removal.
Figure 58: Exhaust Tube Removal.
Battery Bridge
Board
Cooling Fan
Wire Harness
Compressor Power
Wire Harness
Screw