Personal Ambulatory Oxygen System Provider Technical Manual
5) Connect the Output tubing to the Top port on the Product Tank and secure with cable ties. See Figure 48.
Add cable tie to both
top and bottom parts
Figure 48: Connect output tubing
Situation# 2: Removing the New Style Product Tank (PN:4378-SEQ)
1) Follow the steps to "Remove and Replace the Front Cover Assy" on page 40.
2) Remove the Front Cover Assy completely; outlet tubing and front membrane ribbon cable. Set aside.
3) Locate the metal Product Tank and cut the cable ties from the ATF to the Product Tank(Bottom), and O2 Out from the outlet
connector(Top). Remove tubing from ATF to Lower Product Tank barb and set aside for installation. See Figure 49.
Important: Cap off the ATF hose barb with a black hose barb cap.
Cut here
Cut here
Figure 49: Cut Cable Ties
74