Adjustment And Replacement Of The Turn Plate - Siemens MOBILETT Plus HP Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

Mechanical systems

Adjustment and replacement of the turn plate

DANGER
Equipment required: 2 pieces of wood (50 x 70 x 500 mm)
The turn plate mechanism must be removed for adjustment or replacement.
Ensure compliance with the safety information in the section:
"Safety information and protective measures".
Fig. 13
Remove the supporting wheels.
Jack up and block the unit securely 5 cm above the floor. This must be done properly so
that the unit does not tip over.
Remove battery charger D201 and motor power unit D102 (see section "7" and "11").
Disconnect all cables, ground connection and K68 according to (Fig. 13).
Siemens AG
Medical Solutions
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
Shaft
SPR8-220.061.03
Rev. 03
Locking
ring
K68
Page 11 of 28
09.03
CS PS 24
10 - 11
10
Locking ring
MOBILETT Plus HP