8 - 4
Fig. 4
2
3
Fig. 6
Remove the three cable ties (pos. 1 / Fig. 4).
Remove the cable winder assembly.
Remove the two cable ties (pos. 1 / Fig. 5).
Remove the four screws and four nuts (pos. 2 / Fig. 5) and Y-plate (pos. 3 / Fig. 5).
Disconnect the cable from upper plate (pos. 1 / Fig. 6).
Remove the upper plate and distance ring (pos. 2 / Fig. 6).
Loosen the brake carefully to loosen the tension of the cable winder spring.
Remove the securing screw (pos. 1 / Fig. 8)
Pull out the cable winder.
MOBILETT Plus HP
3
1
Fig. 5
1
3
SPR8-220.061.03
Rev. 03
09.03
Miscellaneous
1
2
4
Page 4 of 8
CS PS 24
Medical Solutions