Mechanical systems
Fig. 19

Replacement of the support arm cable harness

DANGER
Equipment required: Rope
Position the unit with the single tank resting on a table. Secure the support arm with a
piece of rope (Fig. 19).
Remove covers "A" and "B".
Remove collimator cover "C".
Siemens AG
Medical Solutions
A
C
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
SPR8-220.061.03
Rev. 03
B
Page 17 of 28
09.03
CS PS 24
10 - 17
MOB00439
10