Mechanical systems
Voltage divider
HT-
Fig. 18
•
Reconnect all electrical connections according to Fig. 18.
•
Install the collimator and replace the cable ties according to the information in chapter 9.
•
Tighten the adjustment screw (Fig. 16) so that the single tank is balanced.
•
Set switch S10 in the OFF position.
Siemens AG
Medical Solutions
Filament transformer
VT
BU
BK
WH
BK
BK
YE RD
GN
OG
Temperature sensor
HT transformer
SPR8-220.061.03
Rev. 03
BK
Stator
for the
rotation
BU
RD
WH
BN
Resistor for
mA measuring
(mA pos)
BK
BK
Page 15 of 28
09.03
CS PS 24
10 - 15
K71
Voltage
divider HT+
BN
BK
OG
GN/YE
BK
BK = Black
BU = Blue
RD = Red
YE = Yellow
GN/YE = Green/ yellow
WH = White
VT = Violet
OG = Orange
BN = Brown
MOBILETT Plus HP