2 - 8
Dis-
played
messages
Err 16
Err 17
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus HP
Mode of
Cause
operation
Generator ON
The capacitors are
charging too rapidly
(> 28 V/s)
Generator ON
The capacitors are
charging too slowly
(< 1.4 V/s)
SPR8-220.061.03
Rev. 03
Display messages
Possible reasons and advice
Check the capacitance in the
capacitor bank (see section "P03 -
Capacitor bank and charging test").
If the capacitance is too low, the
capacitors are disconnected.
If there is an internal error in the
capacitors, the whole row is
disconnected by a fuse (see page
125** and section "Battery circuit
and power transmission unit").
If the error occurs after working on
the capacitor bank, the internal
residual voltage in the modules
may have been too high when they
were connected to each other. This
may have caused the fuse to blow
(see page 125**).
Check that S10 on D7 is switched
off.
If the error occurs after working on
the capacitor bank, check the
cabling and the connections to the
capacitor bank (see page 125**).
Check fuse U8 on D8 and fuse U1
on D7.
Check the capacitance in the
capacitor bank (see section "P03 -
Capacitor bank and charging test").
Internal short-circuiting in a
capacitor or in the capacitor bank
(see section "Battery circuit and
power transmission unit").
Page 8 of 22
09.03
CS PS 24
Siemens AG
Medical Solutions