Replacement and repairs on the collimator
cable tie left side
E
E
Fig. 8
Disconnect the cable to the connection terminal board "D" (Fig. 8).
Disconnect K71 from board D70 (Fig. 8).
Remove the cable ties on the left and the right side (Fig. 8).
Remove the five cable ties according to "F" (Fig. 9).
F
F
Fig. 9