Removal Of Diaphragm-Plate Assembly - Siemens MAMMOMAT 3000 Modular Modification Instructions

Installation of the option w/rh anode/filter combination
Table of Contents

Advertisement

2 - 2
2.

Removal of diaphragm-plate assembly

Fig 2.3
1.
2.
Fig. 2.4
3.
MAMMOMAT 3000 Modular
Service
Lift the holder (4) and carefully push it inside the collimator.
6
Loosen the diaphragm-plate assembly (5) by unscrewing four Allen screws (6).
To release the diaphragm-plate assembly, cut the cable ties on each side of the
tube head
.
Lower the diaphragm-plate assembly and let it hang in the cables (7).
Register 5
5
6
SPB7-230.092.11
Rev. 01
04.98
Disassembly
7
Page 2 of 4
Siemens-Elema AB
SPS-UD
2
Solna, Sweden