P15 - Upper Mas Limit; P16 - Show The Exposure Counter - Siemens MOBILETT Plus HP Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

5 - 8
Pressing the lamp button stores the currently displayed values and also exits the program
P14.
NOTE

P15 - Upper mAs limit

The program enables you to set the upper mAs limit. The available range is
0.5 - 360 mAs in battery operation. The maximum mAs value depends on the actual kV
but can be additionally limited. When selecting an mAs limit value, all prior higher mAs val-
ues will be blocked and vice versa.
To change or check the highest available mAs, activate the program with the lamp button.
The display shows the selected mAs limit, change with mAs + or mAs - buttons.
Example:
kV
SEt 100 = The mAs value is limited to 100 mAs
Press the lamp button to store the currently displayed values and to exit P15.

P16 - Show the exposure counter

This program is activated by the lamp button and shows the actual number of exposures.
Example: 1
kV
599 = The number of exposures is 599.
Example: 2
kV
1
Exit the program by pressing the lamp button.
NOTICE
MOBILETT Plus HP
Always check that the values selected correspond to the desired
values after exiting the service program.
mAs
mAs
mAs
15055 = The number of exposures is 115.055
The exposure counter can not be cleared via any of the service
programs. The value is stored in the CMOS memory.
If the backup battery is depleted, the value could be lost.