Replacing The Vertical Friction Linings - Siemens MOBILETT Plus HP Service Instructions Manual

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

10 - 26

Replacing the vertical friction linings

DANGER
Equipment required: Large caliper, rope, ceiling support.
Raise the single tank to maximum height and secure it with a ceiling support (Fig. 28).
CAUTION
NOTE
MOBILETT Plus HP
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
The ceiling suspension must be able to hold a weight of 100 kg
since the entire weight of the arm will be on the ceiling support
once the spring is removed.
If not, the arm and single tank will fall.
Make sure the ceiling support is sufficient.
Do not stand under the single tank or the arm system during the
work that follows.
SPR8-220.061.03
Rev. 03
09.03
Mechanical systems
Min. 2400
Fig. 28
Page 26 of 28
CS PS 24
MOB00484