Siemens MOBILETT Plus HP Service Instructions Manual page 68

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

8 - 2
Switch on the unit and check that no messages are displayed. If an Err or CAL message
is displayed, proceed as follows:
Change to service mode by switching ON S 1:1 on D1.
Perform the adaptation according to section "P10 - Fast adaptation".
Set the selected mAs limit, refer to the section "P15 - Upper mAs limit".
Set the selected kV step and limit, refer to the section "P14 - Number of kV steps and
upper kV limit".
Read out the exposure counter, refer to the section "P16 - Show the exposure counter".
Switch off S 1:1 on D1 (OFF).
Reattach the system upper cover.
Make an exposure at 90 kV, 10 mAs to check that everything is functioning correctly.
MOBILETT Plus HP
SPR8-220.061.03
Page 2 of 8
Rev. 03
09.03
CS PS 24
Miscellaneous
Siemens AG
Medical Solutions