DAP measuring system (option)
Recalibration of DAP measuring system
This chapter shows how to recalibrate the DAP measuring system by using an external
ionization chamber (dose meter).
Required equipment
•
Dose meter
•
Calibration tool for DAP measuring system
Recalibration procedure
•
System OFF.
•
Disconnect the DAP ionization chamber connector which is connected to board D70
(pos. 1 / Fig. 1).
•
Connect the DAP calibration tool between the connector and board D70.
NOTE
•
System ON.
•
Set up the MOBILETT and the dose meter as shown in Fig. 2.
NOTICE
Siemens AG
Medical Solutions
Fig. 1 Connector
It is irrelevant which connector is the input or the output of the
DAP calibration tool.
Measurement errors.
If a semiconductor detector is used, the temperature/pressure
factor of the dose meter must be considered.
Otherwise a deviation against the factory calibration may occur.
SPR8-220.061.03
Rev. 03
1
Page 1 of 4
09.03
CS PS 24
12 - 1
12
12
12
12
MOBILETT Plus HP