PROCEDURE
• Pack all of the disassembled parts in a sterile barrier sys­
tem in accordance with DIN EN 11607 (e.g., foil­paper
packaging).
Maximum sterilisation temperature and holding time:
121 °C, at least 20 min. or
132 ºC / 134 ºC, at least 3 min.
VALIDATION
The method has been validated according to DIN EN ISO
17665­1 using:
– Euro­Selectomat 666 from MMM (fractionated pre­vacuum
method)
– Varioklav 400 E from H+P Labortechnik (gravitation
method)

5.6 Visual inspection

Inspect all product components after each cleaning, disinfec­
tion and sterilisation. Replace any broken, misshapen or seri­
ously discoloured parts.

5.7 Storage

Store the nebuliser and the accessories used in a dry, dust­
free location equipped with protection against contamination. It
is recommended to use sterile packaging.
– 60 –