Using a standard thermal disinfector for baby
bottles ﴾not a microwave oven﴿
For effective disinfection, use a disinfector with a runtime of at
least 6 minutes. Regarding disinfection, the duration of the dis­
infection procedure and the quantity of water required for this,
follow the instructions for use of the disinfector you are using.
CAUTION
Inadequate disinfection encourages the growth of bacteria and
thus increases the risk of infection. Thorough disinfection has
not been completed until the disinfector automatically switches
itself off or the minimum disinfection time specified in the in­
structions for use of the disinfector has elapsed. Therefore, do
not switch the device off prematurely. Also make sure that the
disinfector is kept clean and check that it is working perfectly
regularly.

4.5 Visual inspection

Inspect all product components after each cleaning and disin­
fection. Replace any broken, misshapen or seriously discol­
oured parts.

4.6 Drying and storage

• You can remove excess water more quickly by shaking all
parts. Remember to hold the cap of the nebuliser upper sec­
tion, otherwise the hinge may be damaged by shaking.
• Place all parts on a dry, clean and absorbent surface and let
them dry completely.
• For drying, between applications, and in particular during
long breaks, wrap the nebuliser in a clean, lint­free cloth
(e.g. a tea­towel) and keep in a dry, dust­free environment.
• During storage, make sure the nebuliser and accessories
are protected from continuous exposure to direct sunlight.