PROCEDURE
• Place all the parts in the cleaning solution.
• Clean the parts thoroughly with a brush.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and ab
sorbent surface.
VALIDATION
The method has been validated using:
– a pHneutral enzymatic cleaner, Korsolex
0.5% (Bode)
– approx. 50 °C warm water
– Duration of cleaning procedure: approx. 5 minutes
Manual disinfection
For manual disinfection, select a suitable instrument disinfect
ant that has been proven to be effective and compatible with
the cleaning agent used.
EQUIPMENT
– Disinfectant
– Water
CAUTION
PARI product components are not suitable for treatment with
instrument disinfectants based on quaternary ammonium
compounds. Quaternary ammonium compounds tend to be
come concentrated in plastic materials, which can cause in
compatibility reactions in patients.
– 58 –
®
Endo Cleaner