VALIDATION
The method has been validated in Europe using:
– G7836 CD cleanerdisinfector produced by Miele (pro
gramme Vario TD
– Standard baskets produced by Miele for nebulisers and ac
cessories (without the connection tubing) or special baskets
from Miele for nebulisers, accessories and the connection
tubing
– Neodisher Mediclean forte 0.5% (alkaline cleaning agent)
and Neodisher Z (citric acid for neutralisation) from
Dr. Weigert.
1) Cleaning: 50 minutes at max. 65 °C; disinfection: 10 minutes at max. 93 °C
DRYING
Even if the cleanerdisinfector is equipped with a drying
function, make sure that no residual moisture is left in the
components.
• If necessary, remove any residual moisture by shaking the
parts and allow them to dry completely.
• Allow the compressor or the PARI CENTRAL to continue
blowing air through the tubing until all of the condensate
has been blown out.
Manual cleaning
As an alternative to mechanical cleaning and disinfection, the
following manual method can be used:
EQUIPMENT
– Cleaning agent
– Brush
– Warm water