2.5 Operating life

Replace the nebuliser after a year at the latest as changes can
occur in the material of the nebuliser over time. This might im­
pair the effectiveness of the treatment.
3

INHALATION

CAUTION
Risk of cross­contamination in the case of a change in pa­
tients (in the professional environment): If the nebuliser is
used for more than one patient, there is a risk that the germs
can be transmitted from one patient to the next via the
nebuliser and its accessories. To prevent this, the nebuliser
and accessories must be cleaned, disinfected and sterilised,
and the connection tubing must be cleaned and disinfected
mechanically or replaced before every change of patient [see
also:  5.1 General].

3.1 Assembling the nebuliser

CAUTION
Check all nebuliser components and the accessories before
each use. Replace any broken, misshapen or seriously discol­
oured parts. Please also follow the instructions for assembly
given below. Damaged components and/or an incorrectly as­
sembled nebuliser may impair functioning of the nebuliser
and thus treatment as well.