3.4
Connecting the 10-port valve
942 Extension Module Vario LQH (2.942.0070)
Connecting the 10-port valve
NOTICE
Connect all capillaries using long PEEK pressure screws (6.2744.090):
1 Connection to the transfer tubing
Connect a capillary with a long pressure screw to the central con-
■
nection of the 10-port valve.
Fasten the other end of the capillary to the coupling on the trans-
■
fer tubing.
2 Connecting the transfer tubing and Dosino
Fasten the capillary to the coupling of the transfer tubing.
■
Connect the other end of the capillary to the Dosino (see manual
■
for the Dosino).
3 Remaining capillary connections
The remaining 10 connections can be used as needed for:
Connecting to the injection valve in the Extension Module.
■
Connecting to the injection valve in the IC instrument.
■
Connecting to different solutions.
■
Connecting to an ion exchanger.
■
etc.
■
3 Installation
■■■■■■■■
17