5.2 Peristaltic pump
1
Tubing olive
3
Filter screw
■■■■■■■■
30
Replacing the filter
Figure 11
Pump tubing connection – Replacing the filter
1 Unscrewing the filter screw
Unscrew the filter screw (11-3) from the tubing olive (11-1) using
the two adjustable wrenches.
2 Replacing the filter
Remove the old filter (11-2) using tweezers.
Use tweezers to place the new filter (11-2) on the filter screw
(11-3) so that it is flat and press it firmly into place with the rear
of the tweezers.
3 Installing the filter screw
Screw the filter screw (11-3) back into the tubing olive (11-1) and
start by tightening it by hand. Finish tightening it using the two
adjustable wrenches.
2
Filter (6.2821.130)
Pack contains 10 pieces.