3.3 Base tray and bottle holder
1
■■■■■■■■
16
2
Figure 8
Mounting the bottle holder
1 Mount the bottle holder onto the topmost instrument in such a way
that the openings in the bottle holder exactly match the screw
threads on the top surface of the instrument.
2 Insert the 2 cylinder screws and tighten them with the 3 mm hex key.
3 Insert both covering stoppers.
After attaching the bottle holder, restore all connections that were loos-
ened at the beginning of the process.
Restoring the loosened connections
1 Plug in all necessary USB cables.
2 Plug in all necessary MSB cables.
3 Plug in the power cord.
4 Mount the drainage tubing again (see manual of the IC instrument).
A longer section of silicone tubing (6.1816.020) may have to be cut
to size and mounted (see also the manual for the IC instrument).
5 If one of the instruments in the stack is equipped with a leak sensor
connection socket, connect the leak sensor (see manual of the IC
instrument).
6 Restore any capillary connections that may have been removed.
■■■■■■■■■■■■■■■■■■■■■■
3