Power Meter Servo Configuration - Keysight Technologies N5166B CXG User Manual

Signal generators
Hide thumbs Also See for N5166B CXG:
Table of Contents

Advertisement

Optimize Performance
Using the Power Meter Servo

Power Meter Servo Configuration

The following procedure is a basic configuration for using the signal generator's Power Meter Servo
mode.
The configuration described below is one possible setup example.
Consider the limits of your DUT and use caution to protect the DUT from
being exposed to too much power.
1. Connect the equipment as shown in
Figure 4-15
2. Set the signal generator frequency and amplitude to the desired settings for your device.
To get the best results and reduce stress on the DUT, the amplitude offset
should be set to match the approximate gain of the device. For example, if
the DUT is an amplifier with 20 dB of gain, set the amplitude offset to 20
dB.
Press AMPTD > More > Amptd Offset, then enter a value.
3. On the signal generator, press AMPTD > Leveling Control > Power Meter Servo Configuration.
4. Press Channel A Setup and configure the power meter. Refer to
Channels
on page 91.
5. Press Return > Return > Settling Error. Set the Settling Error value.
6. Press Return > Return > Leveling Mode. Select the leveling mode.
Power Meter Once performs the adjustment only at the end of any transition (amplitude or
frequency change). After the adjustment is performed and the power is corrected, no further
adjustments are performed until the next transition.
100
Figure
Power Meter Servo Configuration
Power Sensor
Keysight CXG, EXG, and MXG X-Series Signal Generators User's Guide
4-15.
Directional Coupler
Configuring the Power Sensor
For details on each key,
use key help as described
page 60
on
.

Advertisement

Table of Contents
loading

Table of Contents