2.2 Lid

2.2.1 Lid Release

After the run, properly close the lid of the centrifuge, appearing in the display, "rpm/rcf"(A-1) with the
word, "close" (M1). If the centrifuge lid is closed, the code number of the preset rotor is shown in the
display. Even if the rotor is not inserted. By pressing the key, "lid" (7), the lid of centrifuge can be
released. As soon as the electromagnetic lid is completely released, the word, "open" (M2) appears.
The lid of the centrifuge is now able to be opened.
For all number marked text, please refer to Figure 16 .
Figure 16

2.2.2 Lid Lock

The centrifuge lid must be pressed down, until the latch audibly locks. At the same time disappears
the word „open" (M2). As a sign that the centrifuge is ready for starting, appearing in the display,
"rpm/rcf"(A-1), the word "close" (M1). Simultaneously, the word "rotor" (M3) appears, as well as the
code number of the rotor, which is pre-set, i. e. "nr 22x.xx" (M4), along with all rotor specific data, for
example: max. speed, acceleration etc., are available.
For all number marked text, please refer to Figure 16.
ATTENTION: Don't grip your fingers between lid and device or locking mechanism,
© Hermle Labortechnik GmbH
when closing the lid!
Z207MK_V1.21_eng
OPERATION
15