Problem
⑥Oil does not circulate, or the oil
Abnormal sound is
generated.
hole of the cover is clogged.
⑦ Components inside the pump
① Continuous operation at high
Pump surfaces are
extremely hot.
evacuation pressure.
②The specified amount of oil has
not been added. (If the oil
amount is not sufficient, the
cooling effect
of the pump will be reduced.)
③The temperature of the evacuated
gas is high.
④Oil does not circulate, or the oil
hole of the cover is clogged.
①The pump is been filled in excess
A lot of oil splashes
of the specified amount.
out from the outlet
port.
①Deterioration of the O-ring and
The oil leaks
the oil seal of the case and cover.
outside the pump.
Cause
have burnt out.
Continuous
operation
performed at a high evacuation
pressure.
-30-
Measures
⑥Overhaul and clean the oil hole.
⑦Overhaul (replace the damaged
components).
①When it runs by high inhalation
pressure consecutively, there is
no problem in particular in
pump surface temperature to
less than 40K.
②Add the specified amount of
oil.
③Mount cooling equipment such
as a gas cooler at the inlet side.
④Overhaul and clean the oil hole.
①Discharge the oil until it reduces
to the specified amount.
②Install an oil mist trap at the
is
outlet side.
①Check and replace the O-ring
and oil seal.
Reference
6.2
6.2
4.1
6.3
--
6.2
6.3
4.7
4.8
6.2