Failure message
Problem with centrifuge
E-28
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
E-29
The centrifuge cannot be operated.
The run does not start.
E-30
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
E-33
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
E-34-E-36
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
E-40
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
E-41-E-74
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
When to contact Customer Service
If you need to contact customer service, please provide the order no. and the serial no. of your device.
This information can be found on the back near the inlet for the power supply cable.
To identify the software version, proceed as follows:
1. Hold down any of the keys and then switch on the centrifuge.
You enter the system menu.
2. Press the START key
Thermo Scientific
Possible causes and cures
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Check wether you selected the right bucket.
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Overpressure in the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
The centrifuge accelerates too slowly.
Is the rotor properly installed?
Check wether you selected the right bucket.
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
.
7
Trouble Shooting
When to contact Customer Service
Sorvall Legend XF / XFR
7-5