4
Operation
Entering Parameters
6. If available close the rotor with the rotor lid.
Be sure to check all sealings before starting any aeorsol-tight applications.
See the information in the rotor instruction manual.
7. The display shows:
Entering Parameters
Acceleration curve
The Sorvall Legend XF / XFR offers you a total of 9 acceleration and 10 braking curves with which
samples and gradients can be centrifuged.
After the centrifuge is turned on, the last running profile selected is shown.
1. Press the ACC DEC key
The display shows the message "Set acceleration".
2. Select the desired profile by entering a digit between 1-9 on the numeric pad. Curve number 1 is
the slowest and curve number 9 the fastest.
3. Confirm your entry by pressing ENTER
Braking curve
1. Press the ACC DEC key
The display shows the message "Set deceleration"
2. Select the desired curve by entering a digit between 1-9 on the numeric pad. Curve number 1 is
the slowest and curve number 10 the fastest.
3. Confirm your entry by pressing ENTER
Preselecting Speed / RCF
1. Press the SPEED key
The display shows the "RPM" or the "RCF"-value depending on the display setting. Press the
4-4 Sorvall Legend XF / XFR
once in order to open the acceleration profile selection menu.
.
twice in order to open the braking curve selection menu.
.
.
Thermo Scientific