Thermo Scientific Sorvall Legend XF Instruction Manual page 25

Hide thumbs Also See for Sorvall Legend XF:
Table of Contents

Advertisement

TOGGLE key
2. Enter the desired value using the numeric pad.
The digits show in sequential order.
3. Confirm your entry by pressing ENTER
Your entry will be automatically confirmed if you do not press any key for 5 seconds.
Note If an extremely low RCF value has been selected, it will be corrected automatically if the
resulting speed is less than 300 rpm. This is because 300 rpm is the lowest selectable speed.
The relative centrifugal force is given as a multiple of the force of gravity g. It is a unitless numerical
value which is used to compare the separation or sedimentation capacity of various devices, since it
is independent of the type of device. Only the centrifuging radius and the speed come into play in
it:
RCF
r = centrifuging radius in cm
n = Rotational speed in rpm
The maximum RCF value is related to the maximum radius of the tube opening.
Remember that this value is reduced depending on the tubes and adapters used.
This can be accounted for in the calculation above if required.
Running time preselection
1. Press the TIME key
The display shows the message "Set Time".
Enter the desired runtime in hh.mm
Thermo Scientific
to toggle between the two modes.
Explanation of RCF Value
2
n
×
×
------------ -
11.18
=
1000
in order to open the runtime selection menu.
.
r
4
Operation
Entering Parameters
Sorvall Legend XF / XFR
4-5