Thermo Scientific Sorvall Legend XF Instruction Manual page 29

Hide thumbs Also See for Sorvall Legend XF:
Table of Contents

Advertisement

The default value is marked accordingly. You can either confirm the default value by pressing
ENTER
2. Enter the radius using the numeric pad.
The digits show in sequential order.
3. Press ENTER
Setting up a program
1. Enter the program parameters.
2. As described above, please enter braking and acceleration profile, speed/ RCF value, running time,
temperatures, bucket code and bucket radius.
3. Press any of the program store keys for 5 seconds.
Configuring, storing and protecting a program
Programs can be stored and retrieved later by means of the assigned number.
1. Enter the program parameters as described above.
2. Press the program memory key to load any of the programs 1-5.
Press the DIRECTORY key
numeric pad.
Enter the program number and confirm with ENTER key
3. You may want to enter a passcode to protect the data.
After four seconds the display shows the message "Program saved" or "Program locked".
Overwriting a passcode-protected program
If you wish to overwrite a passcode-protected program proceed as follows:
1. Select the program you wish to change.
2. Press the program memory key to load any of the programs 1-5.
Press the DIRECTORY key
numeric pad.
Thermo Scientific
or enter a different radius.
to confirm the new radius.
in order to load any of the programs 1-99 by means of the
in order to load any of the programs 1-99 by means of the