12. Disposable electrodes must be used during defibrillation.
13. Electrodes of dissimilar metals should not be used; otherwise it may cause a high
polarization voltage.
14. The disposable electrodes can only be used for one time.
15. Do not touch the patient, bed, table or the equipment while using the ECG together
with a defibrillator.
16. Do not touch accessible parts of electrical equipment and the patient simultaneously.
17. The use of equipment that applies high frequency voltages to the patient (including
electrosurgical equipment and some respiration transducers) is not supported and
may produce undesired results. Disconnect the patient data cable from the
electrocardiograph, or detach the leads from the patient prior to performing any
procedure that uses high frequency surgical equipment.
18. If WIFI technology is used, in order to maintain compliance with the FCC RF
exposure guidelines, WIFI should be installed and operated with a minimum distance
of 20cm between the radiator and the human body. There should be no shield in or
around the room where WIFI is used.
19. Fix attention on the examination to avoid missing important ECG waves.
20. SHOCK HAZARD - Don't connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer.
21. SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
22. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1 approved to the electrocardiograph.
The operation or use of non-approved equipment or accessories with the
electrocardiograph is not tested or supported, and electrocardiograph operation and
safety are not guaranteed.
WARNING
- 3 -
Safety Guidance