SE-601 Series Electrocardiograph User Manual
Minimum Amplitude:
Noise:
Multichannel crosstalk
AC Filter:
DFT Filter:
Filter
EMG Filter:
LOWPASS Filter:
CMRR
Sampling Frequency
Pacemaker Detection
Amplitude
Width
Sampling Frequency
External Input/Output (Configurable)
Input
Output
WIFI (Configurable)
20 μVp-p
≤12.5µVp-p
≤0.5mm
On /Off
SE-601A:
0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz
SE-601B/C:
0.01Hz/0.05Hz/0.15Hz/0.25Hz/0.32Hz/0.5Hz/0.67Hz
25Hz/35Hz/45Hz/OFF
SE-601A:
150Hz/100Hz/75Hz
SE-601B/C:
300Hz/270Hz/150Hz/100Hz/75Hz
≥115dB (AC OFF)
SE-601A:
≥123dB (AC OFF)
SE-601B/C:
SE-601A:
1,000/sec/channel
SE-601B/C:
16, 000/sec/channel
SE-601A:
± 2 mV to ± 700 mV
SE-601B/C:
± 750uV to ± 700 mV
SE-601A:
0.1ms to 2.0 ms
SE-601B/C:
50μs to 2.0 ms
SE-601A:
10,000/sec/channel
SE-601B/C:
16,000/sec/channel
≥100kΩ; Sensitivity 10mm/V±5%;
Single ended
≤100Ω; Sensitivity 1V/mV±5%;
Single ended