SE-601 Series Electrocardiograph User Manual
NOTE:
1. The order function is only configurable for SE-601B/C.
2. To use the order function, the ECG data management software (DMS) of EDAN must
be installed in the PC.8.1 Loading Orders
Once the electrocardiograph is turned on, you can open the Order Manager screen by pressing
Patient if orders without examination exist on the Order Manager screen.
After you press Examine on the Order Manager screen to return to main screen1, you cannot
open the Order Manager screen by pressing Patient until you complete an order examination
next time.
To load order information:
1. On the electrocardiograph, configure the Order Setup screen.
2. Connect the electrocardiograph to the PC installed with the DMS by using an Ethernet cable
recommended by the manufacturer.
3. Set Server IP, Local IP, Gateway and Subnet Mask in the Transmission Setup window.
For details, please refer to Section 10.5 "Transmission Setup".
4. Select Load on the Order Manager screen to load orders from the DMS.
If Auto Refresh is enabled in Order Setup, the system will automatically refresh the order
information every time you enter the Order Manager screen.
NOTE: If orders are modified on the DMS, the corresponding orders existing on the
Order Manager screen will be refreshed after you load orders from the
software.
- 38 -
Managing Orders