Chroma 19071 User Manual page 32

Hipot tester
Hide thumbs Also See for 19071:
Table of Contents

Advertisement

HIPOT Tester 19071/19072/19073 User's Manual
1.CONTRAST
2.BUZZER VOLUME:
3.EN50191
4.DC 50V AGC :
5.PASS ON
1-15
(2) Use Function Key [F3] to move the highlight to the parameter item you want to set,
then use Function Key [F1] and [F2] to set the parameter data.
(3) System Parameter Setting Table
Item
CONTRAST
BUZZER VOLUME HIGH/MEDIUM/L
EN50191
DC 50V AGC
PASS ON
END OF STEP
EOT
2-8
:
7
HIGH
:
OFF
ON
:
OFF
RMT
LOCK OFST ERR
Range
Default
1∼15
HIGH
OW/OFF
ON/OFF
OFF
ON/OFF
ON
OFF, 0.1sec-
OFF
10sec
ON/OFF
OFF
END OF TEST /
END OF
END OF TIMER
TEST
INC.
DEC.
NEXT
RETURN
7
Adjust the LCD brightness
Adjust the buzzer volume
The output meets to EN50191
requirement (AC Max 3mA, DC Max
5mA). The set and stored
measurement parameters won't be
affected as changing this setting.
Select if open hardware AGC as DC
/ IR voltage setting is 50V~1000V.
(When this item is OFF, hardware
AGC is opened only DC voltage set
to above 1001V automatically.)
When set time to PASS ON, after the
setting time for PASS ON is end then
PASS signal will be deleted
automatically. When set time to
OFF, then PASS signal continue until
press any key to delete.
When the setting is ON, as all
STEPs are end the signal for Under
Test will auto exit from 10mS.
When the setting is OFF, only all
STEPs are end the signal for Under
Test will exit from 10mS.
1. END OF TEST (Pass & Fail &
Under Test signals are activated
after discharging.)
2. END OF TIMER (Pass & Fail &
Under Test signals are activated
before discharging.)
Description