HIPOT Tester 19071/19072/19073 User's Manual
Range
5mA(3~5mA)
3mA(0.3~3mA)
300uA(30~300uA)
30uA(3~30uA)
3uA(0.3~3uA)
300nA(20~300nA)
Note
Select IR suitable current range please follow test voltage and DUT
insulation resistance for counting the quantity of current then follow it to
select suitable current range.
RAMP: Step-up setting voltage needed time, 0 means OFF.
DWELL: Set the required time for test, 0 means OFF. (Do not judge the high and
low limit for insulation resistance during this period.)
FALL : Set the required time for falling from selected voltage to zero, 0 means OFF.
(4) Ground Continue Mode
STEP 1/1
CURR :
0.0A
HIGH :
1.0Ω
DWELL:
0.3s
SELECT MODE
CURR
: Set the output current (0.1A maximum).
: Set the high limit of ground continue (0.1Ω ∼ 5Ω).
HIGH
DWELL : Set the time required for test (0.1 sec ∼ 1 sec). (Do not judge the high
and low limit for ground continue during the time.)
LOW
: Set the low limit for ground continue with the value lower than the high
limit of ground continue or OFF.
(5) Pause Mode
STEP 1/1
PAUSE:PAUSE-MODE
UNDER TEST SIGNAL:OFF
SELECT MODE
2-12
Setting Voltage
50V ~ 250V
0.10MΩ ~ 0.10MΩ
0.10MΩ ~ 0.90MΩ
0.10MΩ ~ 9.00MΩ
0.10MΩ ~ 25.00MΩ
22.0MΩ ~ 90.0MΩ
22.0MΩ ~ 250.0MΩ
0.220GΩ ~ 0.900GΩ
0.200GΩ ~ 2.000GΩ
GC
LOW : OFF
RMT
LOCK OFST ERR
PA
RMT
LOCK OFST ERR
IR Display Value
Setting Voltage
250V ~ 1000V
0.10MΩ ~ 1.00MΩ
0.10MΩ ~ 3.50MΩ
0.10MΩ ~ 25.00MΩ
22.0MΩ ~ 35.0MΩ
0.10MΩ ~ 25.00MΩ
22.0MΩ ~ 250.0MΩ
0.220GΩ ~ 0.350GΩ
25.0MΩ ~ 250.0MΩ
0.220GΩ ~ 2.500GΩ
2.20GΩ ~ 3.33GΩ
0.200GΩ ~ 2.500GΩ
2.20GΩ ~ 50.00GΩ
UP
DOWN
NEXT
EXIT
UP
DOWN
NEXT
EXIT