Descriptions of example
The setting meanings of HEX inputs are as the following:
0xAB = Header
0x01 = Destination Address
0x70 = Source Address
0x02 = Data Field Length
0x23 = Command Code
0x02 = 2:GET (Command Parameter)
0x68 = Checksum
The reading meanings of returning HEX code are as the following:
0xAB = Header
0x70 = Destination Address
0x01 = Source Address
0x02 = Data Field Length
0x7F = Reply Message Command Code
0x00 = Reply Message Return Data
0x0E = Checksum
Offset?
Description:
It queries OFFSET status.
Command code:
0xA3
Parameter:
None
Return data:
1 byte, 0: Off, 1:On or 2:Getting
Example:
Master (0x70): 0xAB 0x01 0x70 0x01 0xA3 0xEB
Slave (0x01): 0xAB 0x70 0x01 0x02 0xA3 0x00 0xEA
Descriptions of example
The setting meanings of HEX inputs are as the following:
0xAB = Header
0x01 = Destination Address
0x70 = Source Address
0x01 = Data Field Length
0xA3 = Command Code
0xEB = Checksum
The reading meanings of returning HEX code are as the following:
0xAB = Header
0x70 = Destination Address
0x01 = Source Address
0x02 = Data Field Length
0xA3 = Command Code
0x00 = 0: Off (Return Data)
0xEA = Checksum
Step Parameters
Description:
It sets all parameters of various steps.
Command code:
0x24
Parameter:
28 bytes
RS485 Interface (For Link Model of 19073 + RS485)