Chroma 19071 User Manual page 20

Hipot tester
Hide thumbs Also See for 19071:
Table of Contents

Advertisement

HIPOT Tester 19071/19072/19073 User's Manual
Changing the Time for Voltage Rising
This instrument contains [RAMP] function. This function is to change the time required for
the outputted test voltage from zero rising to the specified value.
Changing the Time for Voltage Falling
This instrument contains [FALL] function. This function is to change the time required for
the outputted test voltage from the specified value falling to zero.
Current Range Auto Switch
There are two HiPot test AC ranges, one is low AC range 0 ~ 3.000mA, the other is AC
3.01~20mA. There are three DC ranges, low current range 0 ~ 300uA, current range
0.301mA ∼ 3mA and current range 3.01mA ∼ 5.00mA. When the testing current is small,
you can set if to switch the current range to the low one automatically for higher resolution
via software as need.
RS-232 Link (For link model of RS-232 only)
The tester that can use RS-232 interface card and PC mainframe performs link test.
RS-232 Link and RS-485 Interface Multi-Link Test (For link test model of RS-485
only)
The tester that can use RS-232 interface card or install RS-485 interface card and PC
mainframe performs link test. The maximum is 31 sets of instruments to process link test
with PC mainframe as using RS-485 interface card simultaneously. Also, PC mainframe
can control the setting and read the test data. When there is no PC mainframe, multi-link
test can be performed individually but the test data can't be read.
1-2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1907219073

Table of Contents