Pulling In Additional Cables - Leica M500 OH3 User Manual

Hide thumbs Also See for M500 OH3:
Table of Contents

Advertisement

Adjusting focus manually
Warning
Severe injury to tissue can result from manual
adjustment of the working distance!
➩The rotary knob may not be adjusted for manual
adjustment of the working distance while the laser is
in use.
If the focus motor fails, focusing can be set manually with the
rotary knob (1).
1
➩Set the required focus with the rotary knob (1).

Pulling in additional cables

Additional cables can be pulled through the available cable
channels of the Leica M500 OH3 for use with accessories (e.g.
video, DI C500).
Positioning aids (3) are located at the cable channels
(2) and must engage in the counter piece of the cable
channel during installation.
3
➩Loosen cable channels (2) and pull cables in.
➩Secure cables in stretched position of the stand using cable
ties (see arrows).
➩Close cable channels again.
Leica M500 OH3 / Ref. 10 711 894 / Version -
2
3
2
3
Operation
29