Leica M500 OH3 User Manual page 28

Hide thumbs Also See for M500 OH3:
Table of Contents

Advertisement

Operation
Setting focus and zoom speed, focus and zoom position and
eyepiece magnification
Always pay attention to the title line. It displays the
name of the doctor who just saved the settings which
were called up. This is to prevent the settings of other
doctors from being changed or deleted.
➩Select «User settings» in main menu 2 and confirm with «ok».
➩Select «Microscope» under «User settings» and confirm with
«ok».
➩Select either «Focus speed and position», «Zoom speed and
position» or «Eyepiece magnification» under «Function» and
confirm with «ok».
➩Set the desired value with «+/–» and accept with «Save».
Changing contrast and brightness of control device display
➩Go to «Change user» in main menu 1 and confirm with «ok».
A list of names appears for doctors who have saved settings
at the microscope.
➩Select the name of the doctor under «Select user» and
confirm with «ok».
The settings of the selected doctor are active and the dioptre
values, eye-base and eyepiece magnification of the selected
doctor are shown in the display.
➩Continue with «ok».
➩Select «User settings» in main menu 2 and confirm with «ok».
➩Select «Control unit» under «User settings» and confirm with
«ok».
➩Select «Screen contrast/illumination» under «Function» and
confirm with «ok».
➩Select «Screen contrast» or «Background illumination» with
the arrow keys, set the desired value with «+/–» and confirm
with «Save».
Entering or changing user name
➩Go to «Change user» in main menu 1 and confirm with «ok».
A list of names appears for doctors who have saved settings
at the microscope.
➩Select the name of the doctor under «Select user» and
confirm with «ok».
The settings of the selected doctor are active and the dioptre
values, eye-base and eyepiece magnification of the selected
doctor are shown in the display.
➩Continue with «ok».
➩Select «User settings» in main menu 2 and confirm with «ok».
➩Select «Control unit» under «User settings» and confirm with
«ok».
➩Select «User name» under «Functions» and confirm with
«ok».
26
➩Place the flashing cursor under the desired code with <–/–>.
➩Select the code to be set in the code table with «Code +/–»
(code has black background) and set it with «Set/ins».
The code is inserted into the line and the cursor jumps to the
next free position.
➩Search for and insert next code.
If the user name is entered:
➩Select «Done/del».
«Save, continue or cancel» appears on the status line.
➩Select the desired function with the arrow keys.
Changing settings for handgrips (joystick and brakes)
The joystick (1) can be configured for the XY movement of the
microscope or for menu control when using a DI C500.
Setting configuration of joystick:
➩Go to «User settings» in main menu 2 and confirm with «ok».
➩Select «Left handgrip» or «Right handgrip» under «User
settings» and confirm with «ok».
➩Select «X/Y» or «DIC» under «Right handgrip» / «Left
handgrip» and confirm with «Save».
Leica M500 OH3 / Ref. 10 711 894 / Version -
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents