Changing settings for footswitch 1/2
Always pay attention to the title line. It displays the
name of the doctor who just saved the settings which
were called up. This is to prevent the settings of other
doctors from being changed or deleted.
➩Go to «Change user» in main menu 1 and confirm with «ok».
A list of names appears for doctors who have saved settings
at the microscope.
➩Select the name of the doctor under «Select user» and
confirm with «ok».
The settings of the selected doctor are active and the eye-
base, dioptre numbers and eyepiece magnification entered
by the doctor appear in the display.
➩Continue with «ok».
➩Select «User settings» in main menu 2 and confirm with «ok».
➩Select «Foot/handswitch connector 1» or «Foot/handswitch
connector 2» under «User settings» and confirm with «ok».
User configuration:
There are two footswitch configuration variants:
«Standard configuration» and «User configuration».
With the «standard configuration», certain
configurations of the keys and pedals are suggested
and can be saved. With the «User configuration»
variant, the pedals can be configured as desired.
➩Select «User configuration footswitch» under «Function» and
confirm with «ok».
➩Select the desired pedal (❍ becomes G) under «Select
switch» with the right arrow keys under «up/down» in the
sketch in the display.
➩Select the desired function with the left arrow key under
«up/down» (e.g. «focus up»).
➩Confirm with «ok», i.e. upper arrow key (1), and select the
next pedal.
➩When all functions have been selected and confirmed with
«ok», save the setting with «Save», i.e. lower arrow key (2).
The configuration is now saved for the selected doctor.
From a user configuration back to the standard configuration:
➩Select the «Standard configuration foot/handswitch» under
«Function» and confirm with «ok».
➩Accept the recommended configuration with «Save».
The standard configuration of the selected function is
displayed in the sketch (❍ becomes G).
Performing function check on footswitch
A function test can be performed to check the configuration of
the footswitch.
➩Select «User settings» in main menu 2 and confirm with «ok».
➩Select «Foot/handswitch connector 1» or «Foot/handswitch
connector 2» under «User settings».
➩Select «Control foot/handswitch functions» under «Function»
and confirm with «ok».
➩Press any button on the footswitch.
The set function with an arrow symbol
and the ❍ symbol in the symbol picture on the right changes to
G.
Leica M500 OH3 / Ref. 10 711 894 / Version -
Dr. Keller
Select switch using
Working distance long
Working distance short
Zoom up
Zoom down
Additional function 1
Additional function 2
X+
X-
Y+
Y-
Not used
up/down
cancel
ok/save
Operation