➩Press the ALL-FREE button and swivel the microscope carrier
into position B until it engages.
If the mounted accessories (e.g. the tube) do not permit
the 90° swivel movement, turn the tube upward,
perform the swivel movement and then ensure that you
turn the tube back to the working position.
➩Press the B (1) button until the microscope is balanced out.
1
The microscope moves upward:
➩Press the selector switch (2) to the left (–) until the
microscope is balanced out.
2
Leica M500 OH3 / Ref. 10 711 894 / Version -
The microscope moves downward:
➩Press the selector switch (2) to the right (+) until the
microscope is balanced out.
If the microscope cannot be balanced manually, the
accessories are probably not in the balanceable
weight range.
This can be solved for the A/B/C axes by
decreasing/increasing the accessory weight in the
acceptable range (8 kg–11.7 kg of accessories).
The D axis can be adapted by the number of wheel
counterweights (see Page 30).
Operation
2
17